[关键词]
[摘要]
目的 对ADVIA 2120血液分析仪实验动物白细胞分类性能进行评价。方法 将ADVIA 2120血液分析仪常用实验动物白细胞五项分类结果与手工镜检分类结果进行对比,运用SPSS 17.0软件对分类结果进行统计分析。结果 除比格犬、Wistar大鼠、食蟹猴嗜碱粒细胞外,其余各项白细胞分类结果仪器分类和手工镜检相比无显著差异(P>0.05),且相关性较好(r>0.9)。结论 在仪器有白细胞相关提示警告时,无论各项分类指标是否在正常范围,应镜检为宜。
[Key word]
[Abstract]
Objective To evaluate the results of leukocyte classifying counts by ADVIA 2120 Hematology Analyzer. Methods The differential leukocyte counts by ADVIA 2120 Hematology Analyzer were compared with those of manual blood smears under microscopy. The results were analyzed by SPSS 17.0. Results It showed that the relativity of two methods for classification of leukocytes was remarkable (r>0.9) and there was no significant difference (P>0.05) between them, except for the basophils of Beagle’s dogs, Wistar rats, and cynomolgus monkeys. Conclusion The results suggest that when the analyzer warns, it is better to analyze them under microscopy, regardless of whether the classification index in the normal range.
[中图分类号]
[基金项目]